Model AC9900
Measured parameters K+,Na+,Cl-,Ca2+,pH,TCO2
Calculated parameters AG, nCa, TCa
Measurement time 55 seconds
Sample volume 160 μL
Memory 10000 analyzes
Communication interface RS232
Display LCD 240*64 мм
Printer Integral 58 мм
Automatic sample feeding Rotary table with 25 holes

(20 sample wells, 2 for quality control, 1 for emergency analysis, 1 for washing)

Test for air bubbles Yes
Alarm for overflow of the waste container Yes
Power supply AC220V±22V 50Hz 50W
Dimensions 408mmX390mmX230mm
The weight The weight: 7 kg  (without sampling system), 1 kg (sampling system).
Method of measurement Principle of measurement: ion-selective electrode (ISE). TCO2 uses a pressure measurement method.
Environmental conditions Temperature: 10°C~32°C, relative humidity: ≤85%

 

Measured item Measurement scope Resolving power Precision
K+ 0.50-15.00 0.01 CV≤1.0%
Na+ 30.0—200.0 0.1 CV≤1.0%
Cl- 30.0—200.0 0.1 CV≤1.0%
Ca2+ 0.10—5.00 0.01 CV≤1.0%
pH 4.00—9.00 0.01 CV≤1.0%
tCO2 4.0—70.0 0.1 CV≤3.0%